Recall
FDAHealthOctober 9, 2025

Thoratec LLC — Brand Name: HeartMate Product Name: HeartMate II¿ LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, US; 106016, HeartMate II¿ LVAS

⚠️ Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
October 9, 2025

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan.

What's affected?

Product
Brand Name: HeartMate Product Name: HeartMate II¿ LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, US; 106016, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, EU Software Version: N/A Product Description: The HeartMate II LVAS is an axial-flow, rotary ventricular assist system that generates flows up to 10 liters per minute (lpm). One end of the Left Ventricular Assist Device is attached to the apex of the left ventricle; the other end of the device connects to the ascending aorta. The Left Ventricular Assist Device diverts blood from the weakened left ventricle and propels it to the aorta. A small external computer, the System Controller, monitors system operation. A driveline, which passes through the patient s abdomen, connects the implanted pump to the System Controller. The system is powered by a Power Module or Mobile Power Unit (MPU) that is connected to an AC electrical outlet, or by two HeartMate 14 Volt Lithium-Ion batteries. Component: No
Lot code
All

Additional details

Lot Code: All units of the following models are impacted: Model 106015 UDI: 00813024011224 Model 106016 UDI: 00813024011231 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used.

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 248.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.