Recall
FDAHealthOctober 16, 2025

Canon Medical System, USA, INC. — VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.

⚠️ Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.

Is this relevant to me?

Geography
US
Source agency
FDA
Category
Health
Published
October 16, 2025

US distribution to Ohio and Nevada

What's affected?

Product
VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
Lot code
CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Serial Numbers: B7A2462002, B7A2522014.

Additional details

CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Serial Numbers: B7A2462002, B7A2522014.

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 2 units.