American Contract Systems Inc. — Medical convenience kits UDPG88W PICC G-TUBE PACK-230209
⚠️ Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- September 4, 2025
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
What's affected?
- Product
- Medical convenience kits UDPG88W PICC G-TUBE PACK-230209
- Lot code
- 7935611
Additional details
UDPG88W UDI-DI 191072216359 lot 7935611 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected
What should I do?
Recommended action
Class II recall. Voluntary: Firm initiated Quantity affected: 32,433 total.
Get Health recall alerts