Recall
FDAHealthSeptember 4, 2025

American Contract Systems Inc. — Medical convenience kits Item Number/Description UICL44P ADULT CENTRAL LINE INSERTION SUPPLY KIT UICL44R ADULT CENTRAL LINE INSERTION SUPPLY KIT

⚠️ Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Is this relevant to me?

Geography
Nationwide
Source agency
FDA
Category
Health
Published
September 4, 2025

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

What's affected?

Product
Medical convenience kits Item Number/Description UICL44P ADULT CENTRAL LINE INSERTION SUPPLY KIT UICL44R ADULT CENTRAL LINE INSERTION SUPPLY KIT
Lot code
8003911

Additional details

UICL44P UDI-DI 191072194268 Lot 8003911 Bag serial number N/A; UICL44R UDI-DI 191072234018 Lot 8317711 Bag serial number 68244081 * If bag serial number is listed as "n/a", then entire lot is affected

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 32,433 total.