IMPRIS Imaging Inc — IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA
⚠️ There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Is this relevant to me?
- Geography
- Nationwide
- Source agency
- FDA
- Category
- Health
- Published
- September 30, 2025
Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.
What's affected?
- Product
- IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA
- Brand
- IMPRIS Imaging Inc
- Lot code
- IMRIS Part Number(s): 120823-000, UDI/DI 00857534006745
Additional details
IMRIS Part Number(s): 120823-000, UDI/DI 00857534006745
What should I do?
Recommended action
Class I recall. Voluntary: Firm initiated Quantity affected: N/A.
Get Health recall alerts
Delivered to your inbox. Free. No account required.