Recall
FDAHealthSeptember 15, 2025

FUJIFILM Healthcare Americas Corporation — FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33

⚠️ It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.

Is this relevant to me?

Geography
AK, CA, FL, IL, MA, NH, OH, WI
Source agency
FDA
Category
Health
Published
September 15, 2025

Domestic: AK, CA, FL, IL, MA, NH, OH, WI.

What's affected?

Product
FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33
Lot code
Model/Version Number: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33; UDI-DI: (1) 04540217052226, (2) 04540217057436, (3) 04540217057450; Serial Numbers: (1) MP95A8F5A001, MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001, (2) MP95AA783001, (3) MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;

Additional details

Model/Version Number: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33; UDI-DI: (1) 04540217052226, (2) 04540217057436, (3) 04540217057450; Serial Numbers: (1) MP95A8F5A001, MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001, (2) MP95AA783001, (3) MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;

What should I do?

Recommended action

Class II recall. Voluntary: Firm initiated Quantity affected: 16 units.

View original FDA notice →

Get Health recall alerts

Delivered to your inbox. Free. No account required.