SEASPINE ORTHOPEDICS CORPORATION — Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System

Inserter, used as part of a system with an intervertebral body fusion device, may contribute to difficulty disengaging the implant or improper implant positioning, which may may lead to prolonged surgery, intraoperative revision, and may require the use of alternate instrumentation. Devices may also have incorrect UDI-DI marking.

I
ICU Medical Inc. — Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Non-vented stopper included. 100mL, Product Code 21-7002-24

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

I
ICU Medical Inc. — Smiths Medical CADD" Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vented stopper included. 50mL, Product Code 21-7301-24

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

I
ICU Medical Inc. — Smiths Medical CADD" Yellow Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vented stopper included 100mL, Product Code 21-7300-24

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

C
B
Boston Scientific Corporation — Boston Scientific NephroMax Kit, Material Numbers (UPN): 1. M0062101180; 2. M0062101600. (Urology)

The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

H
Haleon US Holdings LLC — Sensodyne PRONAMEL (Potassium nitrate 5%, Sodium fluoride 0.25%) Active SHIELD Toothpaste for Sensitive Teeth, Fresh Mint, Net WT 3.4OZ (96.4g) Tube, 6 tubes x 2 inners, per c

Labeling: Label Mix-up: The outer carton is labelled Fresh Mint. The tube is labelled Cool Mint/Whitening. The toothpaste inside the tube is Fresh Mint as indicated on the outer carton

C
Cardinal Health Inc. — Enoxaparin Sodium Injection, USP, 30 mg/0.3 mL, packaged in 0.3 mL prefilled syringes further packaged in bags of 5 prefilled syringes, Rx Only, Dist. by: Sandoz Inc., Princeton

Correct Labeled Product Mispack: Bags labeled for Enoxaparin Sodium Injection, 80 mg/0.8 mL, contained Enoxaparin Sodium Injection, 30 mg/0.3 mL

P
Philips Medical Systems (Cleveland) Inc — Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Pinnacl

Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.

B
Boston Scientific Corporation — Boston Scientific UroMax Ultra Kit, Material Numbers (UPN): 1. M0062251200, 2. M0062251210, 3. M0062251220, 4. M0062251230, 5. M0062251240, 6. M0062251260,

The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

B
Boston Scientific Corporation — Boston Scientific Encore 26 Inflation Device, Material Numbers (UPN): 1. H74904526011; 2. H74904526052; 3. M0067101140; 4. M001151050; 5. M001151062; 6. M0

The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

B
Boston Scientific Corporation — Boston Scientific Encore Advantage Kit, Material Numbers (UPN): 1. H74904527011; 2. H74904527052. (Interventional Cardiology and Peripheral Interventions)

The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

T
Tandem Diabetes Care, Inc. — t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem

A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

W
WASATCH PHOTONICS — Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.

Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.

H
Hans Kissle Company, LLC — Hans Kissle Red Bliss Potato Salad

Contains undeclared wheat. The top label identifies the product as potato salad, but the unit contains pasta salad.

B
Baxter Healthcare Corporation — Novum IQ Syringe Pump, Product Code 40800BAXUS, infusion pump

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).

A
AVID Medical, Inc. — Halyard, ENT PACK. Model Number: JACK421-04. Convenience kit.

Potential for open header bag seals, compromising sterility.

B
Baxter Healthcare Corporation — Novum IQ LVP (Large Volume Pump), Product Code 40700BAXUS, infusion pump

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).

B
Beckman Coulter Inc. — DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer

On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.

B
Baxter Healthcare Corporation — Novum IQ Syringe Pump, Product Code 40700BAX, infusion pump

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).

A
AVID Medical, Inc. — Halyard, CPT BASIC IR PACK. Model Number: 1646381. Convenience kit.

Potential for open header bag seals, compromising sterility.

M
Merit Medical Systems, Inc. — PhD Hemostasis Valve, REF: IN4802/K, IN4802/T, IN4918/A, IN7802/C, IN8802/C, IN9802, IN9802/A, K05-02166B, K05T-02138A, K12T-10798A, MAP800/B, MAP800/T, MAP801/B, MAP802/

Hemostasis valve has a manufacturing defect and when used the valve cap may remain depressed, leading to potential leakage, which may result in procedure inconvenience and/or hemorrhage.