GE Medical Systems, LLC — LOGIQ P9 R4.5 ultrasound system with system software version R4.5.7., Model Number 5877533

The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.

C
Cardinal Health 200, LLC — Salem Sump PVC Tubes: Product Code/Product Name: 8888264911E ENFit" SALEM SUMP" PVC TUBE 10FR 36IN 8888264929E ENFit" SALEM SUMP" PVC TUBE 12FR 48IN 8888264945E ENFi

Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.

G
GE Medical Systems, LLC — LOGIQ P10 series with software version R4.5.7 Model Number 5877534

The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.

H
HeartSine Technologies Ltd — HeartSine Pad-Pak, REF: PAD-PAK-01, PAD-PAK-02, PAD-PAK-03, PAD-PAK-04, PAD-PAK-07, PAD-PAK-03J, PAD-PAK-04J

Due to customer complaints, defibrillator pads may contain damage pins that disrupt the mechanical fit and electrical contact.

C
Cardinal Health 200, LLC — Salem Sump PVC Tubes: Product Code/Product Name: 266122 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) 266130 Salem Sump" Stomach Tube,

Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.

D
D & M Smoked Fish, Inc. — Haifa COLD SMOKED SALMON 8 OZ.

Contaminated with Listeria monocytogenes

D
D & M Smoked Fish, Inc. — Haifa COLD SMOKED SEABASS 8 OZ

Contaminated with Listeria monocytogenes

B
Baxter Healthcare Corporation — Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400) Thermometers and Welch Allyn SureTemp Plu

Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing fever-related symptoms or infection.

C
CPM Medical Consultants, LLC. — Fuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS200025L, OS200120L, OS200125, OS200125L, OS200135, OS200324,

Instructions for use for a foot plating system is being updated to remove the second option of requirements for sterilization which was 3 minutes exposure time at 134-137-degrees C (273-279-degrees F) and 40mBar for 5-10 minutes drying time, because these requirements do not apply in the United States.

P
Percussionaire Corporation — Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) with In-Line Valve Sentec Instructions for Use

IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure.

E
E
ELEKTA SOLUTIONS AB — MOSAIQ Oncology Information System Software Builds 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 Software Version and UDI Codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201

When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule.

B
Beckman Coulter Inc. — SYNCHRON Systems Direct LDL Cholesterol (LDLD) Reagent, REF: 969706

Beckman Coulter has identified that there is an increased probability Synchron Systems Direct LDL Cholesterol (LDLD REF 969706) reagent lot M404166 may generate false high results. The issue was discovered internally and confirmed on August 19, 2025 by Beckman Coulter.

L
A
ACCORD HEALTHCARE, INC. — Levothyroxine Sodium Tablets, USP, 88 mcg (0.088 mg), 90 Tablets per bottle, Rx Only, Manufactured for Accord Healthcare, Inc., Raleigh, NC 27617, Manufactured by: Intas Phar

Subpotent: During long term stability testing of Levothyroxine Sodium Tablets USP for 88 mcg, the assay content was observed below the approved specification range.

S
Surgical Theater Inc — Surgical Planner, BrainSTORM, software version BST.9.3.0, orthopedic augmented reality

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

S
Surgical Theater Inc — SuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological system

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

S
Surgical Theater Inc — Endo Surgical Navigation Advanced Platform Patient Engagement (EndoSNAP), software version SNP.1.12.0, imaging processing radiological system

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

S
Surgical Theater Inc — SpineAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SPR.1.2.0, orthopedic augmented reality

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

S
Surgical Theater Inc — Surgical Navigation Advanced Platform (SNAP), software version SNP.3.12.0, imaging processing radiological system

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

S
Surgical Theater Inc — Surgical Navigation Advanced Platform Patient Engagement (PE SNAP), software version SNP.3.12.0, imaging processing radiological system

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

S
Spacelabs Healthcare, Ltd. — Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System

Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record

S
Surgical Theater Inc — SyncAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SNP.1.12.0, imaging processing radiological system

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

M
Medtronic Sofamor Danek USA Inc — Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073.

Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.

F
FUJIFILM Healthcare Americas Corporation — FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33

It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.

P
Private Label Skin Care Inc. — edunn clarity, BP Treatment Cleanser 10% (Benzoyl Peroxide 10%), 7 oz (201g) Uptown Medi Spa, Phoenix, AZ.

Chemical Contamination: This recall has been initiated due to elevated levels of benzene that was discovered during post consume awareness with benzoyl peroxide products.

P
Private Label Skin Care Inc. — ARTISAN OF SKIN (Benzoyl Peroxide 10%) , 7 oz (201 g), Artisan of Skin Beverly Hills, CA 90210

Chemical Contamination: This recall has been initiated due to elevated levels of benzene that was discovered during post consume awareness with benzoyl peroxide products.

P
Private Label Skin Care Inc. — mel rx Skin, NO DRAMA Benzoyl Peroxide 10% Acne Treatment, 7 oz, 201 g Mel Skin Rx Beverly Hills, CA 90210

Chemical Contamination: This recall has been initiated due to elevated levels of benzene that was discovered during post consume awareness with benzoyl peroxide products.