ICU Medical, Inc. — Plum Duo Infusion System, List Number 40002-04-01

ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback for touchscreen inputs.

H
Hologic, Inc — Brevera Breast Biopsy System Disposable 9 Gauge Needle. Biopsy Instrument.

Presence of particulates in affected devices that can be deposited into breast tissue during use.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20 OR Table; Model Number: 722023;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20 Biplane; Model Number: 722013;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD10; Model Number: 722026;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20; Model Number: 722006;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

I
Immunotech A.S. — Estrone RIA, REF: DSL8700

The affected Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability. Only in some (very rare) cases may the issue occur.

I
P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD10F; Model Number: 722002;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD10/10; Model Number: 722027;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20; Model Number: 722028;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Product Name: Allura Xper FD10C; Model number: 722001;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD10; Model Number: 722003;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

I
ICU Medical, Inc. — ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

I
ICU Medical, Inc. — ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD10/10; Model Number: 722011;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20/20; Model Number: 722038;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD10/10; Model Number: 722005;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20; Model Number: 722012;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

D
Diagnostica Stago, Inc. — Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection in plasma or serum of the anti-heparin-platelet factor

Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20/10; Model Number: 722029;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

D
D. Coluccio & Sons Inc. — Sapori Amaretti Almonds 175 g Cookies

Contains elevated levels of hydrocyanic acid

M
Medinatura New Mexico, inc. — ClearLife Allergy Nasal Spray, Extra Strength, 0.68 fl. oz. (20 mL) bottles, Distributed by: MediNatura, 10421 Research Rd., SE, Albuquerque, NM 87123, NDC 62795-4006-9,

Microbial Contamination of Non-Sterile Products: The products have been found to contain yeast/mold and microbial contamination identified as Achromobacter.

M
Medinatura New Mexico, inc. — ReBoost Nasal Spray, 0.68 fl. oz. (20 mL) bottles, Distributed by: MediNatura, 10421 Research Rd., SE, Albuquerque, NM 87123, NDC 62795-4005-9, UPC 787647101863

Microbial Contamination of Non-Sterile Products: The products have been found to contain yeast/mold and microbial contamination identified as Achromobacter.

M
Medline Industries, LP — Medline medical procedure convenience kits labeled as: 1) RR-GYN LAPAROSCOPY PACK-LF, Kit SKU DYNJ0395022O; 2) PELVISCOPY PACK-LF, Kit SKU DYNJ0415776Q; 3) DIVA PACK, K

Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Mo

Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.

E
Elekta, Inc. — Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11.1 Product Description: Leksell GammaPlan¿ is a computer-based system design

Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.

M
Medline Industries, LP — Medline medical procedure convenience kits labeled as: 1) PACER / CATH LAB PACK-LF, Kit SKU DYNJ0972604A; 2) PACEMAKER/ICD PACK, Kit SKU DYNJ33822L; 3) VENOUS ACCESS PA

Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.

M
Medline Industries, LP — Medline medical procedure convenience kits labeled as: 1) MEM MAXILLOFACIAL PACK-LF, Kit SKU DYNJ0660040R; 2) PLASTICS VCH, Kit SKU DYNJ39225K; 3) WT BREAST PACK, Kit S

Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.