Recall

All recalls8,436 total

N
S
Stryker Communications — Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034

Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.

W
W
A
American Contract Systems Inc — AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP PACK, MODEL/ITEM NUMBER AMAH40N; 2)TOTAL KNEE DAN MYER, MODEL/ITEM NUMBER C

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

M
MAQUET GMBH — Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.

Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.

B
Baxter Healthcare Corporation — Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS

Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The level of underinfusion is variable based on the current infusion rate, the duration the pump has been running at this flow rate, and the magnitude of the rate change. The longer the duration the pump has been running at the current infusion rate and the larger the magnitude of the rate change, the larger the underinfusion that would be experienced. Additionally, Baxter has identified an increase in customer reports of over and underinfusion potentially due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Consistent with the instructions for use, customers should ensure that: 1) The door is fully open before loading the set. 2) The tubing is taut and loaded without slack in the pumping channel.

A
American Contract Systems Inc — AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON ARRIVAL KIT, Model/Item Number LLBN11C; 2) LABOR & DELIVERY PACK, Model/Item Num

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

B
Baxter Healthcare Corporation — Spectrum IQ Infusion Pump, Product Code 3570009

Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

D
D
D
D
D
D
D
D
A
American Contract Systems Inc — AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) SFCH CRANI PACK (PS 132432), MODEL/ITEM NUMBER UICR10W; 2) SFCH SHUNT PACK (PS 139798),

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

A
American Contract Systems Inc — AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK - 205947, MODEL/ITEM NUMBER ANCV78BC; 2) VASCULAR PACK - 232024, MODEL/ITEM N

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

A
American Contract Systems Inc — AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) A-C SUPPLEMENT PACK - 205946, Model/Item Number ANCV67X; 2) KIDNEY PERFUSION PACK - 208

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

A
American Contract Systems Inc — AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L; 2) LAP CHOLE PACK, Model/Item Number HNLC80J;

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

A
American Contract Systems Inc — AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAPAROSCOPY PACK, MODEL/ITEM NUMBER EVLP61I; 2) PELVISCOPY, MODEL/ITEM NUMBER EVPV29G.

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

A
American Contract Systems Inc — AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) CYSTO, Model/Item Number ASCY24A; 2) CYSTO PACK, Model/Item Number BUCY78E; 3) CYSTO P

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

A
American Contract Systems Inc — AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ROBOT HYST PACK, MODEL/ITEM NUMBER GRRB10E; 2) VAG HYST PACK, MODEL/ITEM NUMBER SPVH62P.

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

P
PIE Medical Imaging B.V. — 3mensio Workstation (Vascular Fenestrated) software

When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage

R
Remel, Inc — Blood Agar,5% Sheep Blood 100/PK, Product Number R01202

The products may contain surface and subsurface contamination of Listeria monocytogenes.

← Previous

Page 110 of 169

Next →