Recall

All recalls7,877 total

A
AVID Medical, Inc. — MAJOR THORACOSCOPY BASIN. Medical convenience kit.

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

A
AVID Medical, Inc. — PARTIAL NEPHRECTOMY PACK. Medical convenience kit.

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

A
AVID Medical, Inc. — HVI CARDIAC ROBOT PACK. Medical convenience kit.

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

C
CooperSurgical, Inc. — Endosee System Convenience Kit. Model Number: ES-TRAY. Endosee System Convenience Kit

The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.

A
AVID Medical, Inc. — GENERAL ROBOTIC PACK. Medical convenience kit.

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

C
CooperSurgical, Inc. — HSG Procedure Tray. Model Number: 6050T. The HSG Procedure Tray is for use in hysterosalpingography and sonohysterography procedures.

The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.

A
AVID Medical, Inc. — DAVINCI UROLOGY. Medical convenience kit.

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

A
AVID Medical, Inc. — GS ENDOSCOPY PACK. Medical convenience kit.

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

A
AVID Medical, Inc. — CT DAVINCI. Medical convenience kit.

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

A
AVID Medical, Inc. — DA VINCI PACK URO/PROSTATE. Medical convenience kit.

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

B
Boston Foundation For Sight — BostonSight SCLERAL Lens

Select orders of lenses were manufactured with errors impacting sagittal height, haptics, or both.

B
Boston Foundation For Sight — BostonSight PROSE Lens.

Select orders of lenses were manufactured with errors impacting sagittal height, haptics, or both.

B
Boston Scientific Corporation — Torflex Transseptal Guiding Sheath Kit (TFK-M) TF85-32-63-45 TORFLEX 8.5F 63CM 45DEG TF85-32-63-55 TORFLEX 8.5F 63CM 55DEG Used for the percutaneous introduc

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

B
Boston Scientific Corporation — VersaCross Access Solution (VXSK), Product Description and ID: VXSK (180J RF WIRE 63CM/45DEG/D0) VXSK0101 VXSK (180J RF WIRE 63CM/45DEG/D1) VXSK0102 VXSK (180J

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

B
Boston Scientific Corporation — VersaCross RF Wire (VXW). Indicated for creation of an atrial septal defect in the heart. Product ID VXW0002 Device approved and commercialized in Japan only.

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

B
Boston Scientific Corporation — VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACROSS SHEATH (63CM 45DEG D0)-J Used for the perc

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

O
Ortho-Clinical Diagnostics, Inc. — Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version: Not Applicab

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

O
Ortho-Clinical Diagnostics, Inc. — Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier II Model/Catalog Number: 8231474 Software Version: N/A Product

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

O
Ortho-Clinical Diagnostics, Inc. — VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides);

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

O
Ortho-Clinical Diagnostics, Inc. — VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides);

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

S
SEDECAL SA — MobileDiagnost wDR 2.2; Model Number REF 9890-010-89522, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — Model Number SM-32HF-B-D-C; 32KW 50G, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — Model Number 40KWFXPLUS.004, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — Model Number SM-40HF-B-D-C; 40KW 55C/60C, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — Model Number SM-40HF-B-D-C; 40KW, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — Model Number SM-40HF-B-D-C; 40KW 70C, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — SM_40HF_B_D_VIR3, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — SM_40HF_BDC3, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — Model Number SM-20HF-Batt; 20KW ANALOG, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — SHIMADZU RF.004, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — Model Number SM-40HF-B-D-C; 40KW.002, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — Model Number 40KW-FX.899, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — Model Number 40KWFXPLUS-710CW.007, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — Model SM-40HF-B-D-C; 40KW - CANON READY, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — Model Number SM-40HF-B-D-C; 40KW.005, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

S
SEDECAL SA — SM_40HF_B_D_VIR3.004, Mobile X-ray system

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

← Previous

Page 110 of 158

Next →