Steris Corporation — HarmonyAIR A-Series Surgical Lighting System

There is a potential for paint delamination (chipping/peeling) on the lighting system's light heads.

S
Steris Corporation — HarmonyAIR A-Series Surgical Lighting System

The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.

S
Steris Corporation — ALYON Surgical Lighting System

The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.

S
Steris Corporation — AMSCO 7052HP Washer/Disinfector Model/Catalog Number: 7052HP Product Description: The AMSCO 7052HP Washer/Disinfector is intended for use in the cleaning and intermediate-lev

Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

S
Steris Corporation — AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the cleaning and intermediate-le

Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

S
Steris Corporation — Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems

The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.

S
Steris Corporation — Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting Systems

The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.

S
Steris Corporation — Barco MNA with the HexaVue IP Integration System, Model/Catalog Number: MNA-6x0-H

This is a sub-recall of event RES 96885. The supplier identified that during their manufacturing process for a limited batch of MNA-6x0 (adapter) components the incorrect configuration file was used.