Recall

philips medical systems nederland b.v.

50 recalls on record

Get alerts
P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Philips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Dete

Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Philips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xpe

Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063, 722221), Azurion 3 M15 (722064, 722222), Azurion 5 M12 (722227), Azurion 5 M20

It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10 OR Table (722022, 722033), Allura Xper FD10

It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12 (722067), Azurion 7 B20 (722068), Azurio

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descriptions and corresponding model numbers. 1. A

Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Integris-Allura system; System Code Description (Model Numbers): Integris CV Cesar-Powerpack-Visub-Nicol (722030), Integris Allura 15 & 12 (monoplane) (722043)

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura system; System Code Description (Model Numbers): Allura Xper FD10C (722001), Allura Xper FD10F (722002), Allura Xper FD10 (722003, 722010, 722026), Allu

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722

Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) intended to be used in combination with a Phili

Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Vue Motion V12. Product Number: 1017979.

Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20/20 OR Table; System Code: 722039;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — ALLURA Xper FD10/10; System Code: (1) 722005, (2)722011, (3)722027;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — ALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20/15; System Code: 722058;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20/20; System Code: 722038;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — ALLURA Xper FD20; System Code: (1) 722006, (2)722012, (3)722028;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 7 M20; System Code: (1)722079, (2)722224, (3)722234, (4)722282 (OUS only);

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD10C; System Code: 722001;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 7 B12; System Code: (1)722067, (2)722225, (3)722235;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — ALLURA Xper FD10; System Code: (1) 722003, (2)722010, (3)722026;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20/15 OR Table; System Code: 722059;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — ALLURA Xper FD20 Biplane; System Code: (1) 722008, (2)722013;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 5 M12; System Code: (1)722227, (2)722231;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — ALLURA Xper FD20 Biplane OR Table; System Code: (1) 722020, (2)722025;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — ALLURA Xper FD10F; System Code: 722002;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20/10; System Code: 722029;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — ALLURA Xper FD20 OR Table; System Code: (1) 722015, (2)722023, (3)722035;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 3 M15; System Code: (1)722064, (2)722222, (3)722230, (4)722280 (OUS Only);

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 5 M20 System Model Numbers: (1)722228, (2)722232, (3)722281(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 3 M12 System Model Numbers: (1) 722063, (2) 722221, (3) 722229; Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 7 B12 System Model Numbers:(1) 722067, (2) 722225, (3) 722235; Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 7 M12 System Model Numbers: (1)722078, (2)722223, (3)722233; Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Azurion 3 M15 System Model Numbers: (1) 722064, (2) 722222, (3) 722230, (4)722280(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD10; Model Number: 722010;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20 Biplane; Model Number: 722008;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20 OR Table; Model Number: 722023;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20 Biplane; Model Number: 722013;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD10; Model Number: 722026;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20; Model Number: 722006;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD10F; Model Number: 722002;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD10/10; Model Number: 722027;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

P
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Allura Xper FD20; Model Number: 722028;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.