Recall

philips medical systems (cleveland) inc

2 recalls on record

Get alerts
P
Philips Medical Systems (Cleveland) Inc — Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Pinnacl

Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.

P
Philips Medical Systems (Cleveland) Inc — Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software

Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,